CVE-2026-105279
J4Care dcm4che Missing Authentication for Critical Function
- Critical 9.3
Description
In the default configuration, the archive accepts DICOM associations from an unregistered calling AE. A remote peer can use C-FIND and C-GET to enumerate and retrieve patient information and medical images. It can also use C-STORE to file fabricated study content under an existing patient identity.
Key facts
- Published
- 16 Oct 2026
- Last changed
- 16 Oct 2026
- Assigned by
- icscert
- EPSS
- 0.274% likely to be exploited in the next 30 days, higher than 20.098% of CVEs (scores for 18 Nov 2026)
- SSVC (CISA)
- Exploitation: none; Automatable: yes; Technical impact: total
Severity scores
Each score is shown as its source gave it. They are not averaged or merged.
- 9.1CVSS 3.1, Criticalfrom the CNA (icscert)
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N - 9.3CVSS 4.0, Criticalfrom the CNA (icscert)
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N
What to do
- J4Care has provided a fix for the vulnerability. J4Care recommends users implement release 5.35.2 or later.CNA solution statement
Historical context and remediation
This reference retains the original CVE publication date. A recent record update does not mean a new attack occurred.
No separately sourced impact summary is available on this page yet. Consult the description, version data and dated exploitation evidence.
Version statuses below come from the CNA. An “unaffected” version is not necessarily a currently supported release or proof of a patch date. EPSS is a dated risk estimate, not a count of historical attacks.
Affected products
| Vendor | Product | Versions |
|---|---|---|
| J4Care | dcm4chee-arc-light |
|
Timeline
- 16 Oct 2026 Published